Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add filters








Language
Year range
1.
Rev. Soc. Bras. Med. Trop ; 51(3): 387-389, Apr.-June 2018. tab, graf
Article in English | LILACS | ID: biblio-1041461

ABSTRACT

Abstract INTRODUCTION: Several factors can cause false-positive results in the galactomannan (GM) test; however, others remain unknown. Presently, the impact of airborne contamination by Aspergillus conidia during enzyme-linked immunosorbent assay (ELISA) remains uninvestigated. METHODS: We studied 12 A. fumigatus isolates. Fungal conidia were serially diluted and tested for GM detection using the Platelia® Aspergillus enzyme immunoassay (EIA). RESULTS: The conidia concentration required for an EIA-positive result was 4.8 × 103 (median). CONCLUSIONS: This is the first study to evaluate the impact of environmental contamination on the Platelia® Aspergillus EIA assay. Only massive contamination can interfere with GM optical readings, suggesting that environmental contamination does not cause false-positive test results.


Subject(s)
Humans , Aspergillus fumigatus/isolation & purification , Spores, Fungal , Enzyme-Linked Immunosorbent Assay/adverse effects , False Positive Reactions , Mannans , Enzyme-Linked Immunosorbent Assay/methods
SELECTION OF CITATIONS
SEARCH DETAIL